CKD HYN Direct Acting 2/3-Port Pinch Valve (for High Purity Fluids)

HYN is a direct acting pinch valve that shuts off the fluid by pinching the tube. It is used for water, pure water and chemical liquids, and the tube is removable. 2-port NO and NC types and a 3-port universal type are available, with compatible tubes of O.D. x I.D. ø3x1, ø5x3 and ø8x6.

CKD HYN Direct Acting 2/3-Port Pinch Valve (for High Purity Fluids)CKD HYN Direct Acting 2/3-Port Pinch Valve (for High Purity Fluids) 2

Specifications

Working FluidWater / pure water / chemical liquids (fluids that do not corrode wetted part materials)
Valve StructureDirect acting pinch valve: 2-port NO (closed when energized) / 2-port NC (open when energized) / 3-port universal
Working Pressure0 to 0.05 MPa (0 to 0.02 MPa for HYN-8)
Fluid Temperature5 to 50 °C
Ambient Temperature0 to 40 °C (no freezing)
Frequency60 cycles/min. or less
Mounting OrientationUnrestricted (avoid vertical mounting with the coil down)
Compatible Tube (silicone tube)HYN-3: ø3 x ø1 | HYN-5: ø5 x ø3 | HYN-8: ø8 x ø6
Rated VoltageAC: 100 V (50/60 Hz) / DC: 12 V, 24 V (voltage fluctuation range ±10%)
Power Consumption (12/24 VDC)HYN-3: starting 15 W, holding 4 W | HYN-5 and HYN-8: starting 30 W, holding 8 W
Max. Current (100 VAC)HYN-3: starting 0.26 A, holding 0.06 A | HYN-5 and HYN-8: starting 0.55 A, holding 0.14 A
Thermal ClassHYN-3: Class 120 (E) / HYN-5 and HYN-8: Class 130 (B)
WeightHYN-3: 0.18 kg | HYN-5: 0.36 kg | HYN-8: 0.37 kg
RatingHYN-3: continuous / HYN-5 and HYN-8: intermittent rating for AC 100 V (max. continuous power ON time within 10 minutes, DUTY ratio one half or less), continuous for DC