The Ecotec E3000 is a multi-gas sniffer leak detector for final testing of refrigerators, freezers, automotive air conditioners and similar products. It measures four gases simultaneously, and the IGS algorithm reduces false alarms caused by background gases such as foaming agents. The sniffer probe handle has a leak rate display and illumination.
| Measuring Method | Multi-gas sniffer leak detection; four different gases measured simultaneously |
| Minimum Detectable Leak Rate | R134a 0.05 g/a (0.002 oz./yr.) / R600a 0.05 g/a (0.002 oz./yr.) / helium 1×10⁻⁶ mbar l/s |
| Measuring Scale | 0.05-999.99 g/a (0.002-99.999 oz./yr.) |
| Sensor Response Time | 0.3 s |
| Response Time Incl. Sniffer Line | 0.8 s |
| Maximum Number of Gases Detectable Simultaneously | 4 |
| Leak Rate Units | g/a, oz./yr., mbar l/s, Pa m³/s, ppm |
| Start-up Time | Less than 2 min |
| Dimensions (W × H × D) | 610 × 370 × 265 mm (24 × 14.6 × 10.4 in.) |
| Weight | 34 kg (75 lb) |
| Gas Flow | 160 sccm |
| Ambient Temperature Range | 10-45 °C (50-113 °F) |
| Ordering Versions | 230 V, 50 Hz / 100/115 V, 50/60 Hz; RC version also available |
| Interfering Gas Suppression | IGS (interfering gas suppression) algorithm ignores background gases such as foaming agents and distinguishes R600a from cyclopentane and isopentane |
| Function Check | Function check and calibration at any time with the built-in ECO-Check reference leak |
| Certification | cTÜVus Certified (NRTL checked) |